33477 Solar Sensor ISO 17025 Accredited Calibration to ASTM G207 with Documentation

Visão Geral

The 33477 calibration service is performed indoors using a reference pyranometer of the same make and model. Reference pyranometers are traceable to the World Radiometric Reference (WRR). This calibration is ISO 17025 accredited to ASTM G207 standard.

imagens

Compatibilidade

The 33477 calibration service is suitable for the following solar sensors:

  • Hukseflux SR20-T1, SR20-T2, SR20-T1-TR, SR20-D2, SR30-D1, SR05-DA1, SR05-DA2, SR11, SR03, and LP02 
  • Kipp & Zonen CMP11, CMP10, SMP11, SMP10, CMP6, and CMP3

Perguntas Frequentes Relacionadas

Number of FAQs related to 33477: 1

  1. We offer three pyranometer calibration services:

    • 33477 ASTM G207 is an indoor calibration. The reference must be the exact same make and model as the sensor being calibrated. For example, a CMP11 must be used to calibrate a CMP11. Silicon detector sensors such as the CS300 are calibrated this way. This is the same method that the manufacturers of ISO 9060 sensors use to calibrate. Good.
    • 33476 ASTM E824 is a more rigorous outdoor calibration. The reference must be the same classification or ISO 9060 specification—for example, secondary standard or second class. Better.
    • 33475 ASTM G167 is an outdoor calibration. The reference is a transfer standard to WMO that has been yearly calibrated to a primary standard. Best.